126 Troubleshooting and Maintenance Manual
5 CI Maintenance
To Clean the Reagent Gas Supply Lines
Materials needed
Clean, dry nitrogen
Heat gun
Tubing cutter (8710-1709)
Procedure
If the reagent gas lines become contaminated, they can be cleaned.
1 Disconnect the reagent gas tubing from the gas supply, the gas purifier,
and the MS.
2 Cap the gas purifier following the instructions on the label.
3 Connect one end of the tubing to a supply of clean, dry nitrogen and turn
on gas flow.
4 Use the heat gun to warm the tubing, starting at the supply end and
working your way to the free end.
5 Repeat for any other pieces of tubing that need to be cleaned.
6 Reconnect the tubing to the gas supply, gas purifier, and MS. Follow the
instructions on the gas purifier label.
WARNING
Do not heat the gas tubing when reagent gas is flowing.
CAUTION
Do not put liquids into the tubing. Do not heat the tubing when it is connected to the
MS.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals