176 Troubleshooting and Maintenance Manual
8 Electronics
Bus board
This board is mounted on the bottom of the electronics module. It acts as a bus
connecting the two main boards.
PLX card
This card, located inside the SmartCard module, provides fast
communications between the processor module and other parts of the MS
system. It speeds up the entire system by removing communications tasks
from the central processor.
LAN/MS Control Card (Smartcard 4)
The LAN/MS control card is located to the left of Main Boards 1 and 2 on the
electronics panel. The LAN/MS control card has two main functions:
Provide a communication interface between the MS and the data system
Provide real-time control of the MS, freeing the data system for other tasks
Functional areas of the LAN/MS control card include:
Instrument controller
Data processor
Main processor
Serial communication processor
Network communication controller
Remote start processor
Random access memory (RAM)
Status LEDs
LEDs on the LAN/MS control card are visible near the LAN connector on the
rear panel.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals