58 Troubleshooting and Maintenance Manual
3 CI Troubleshooting
Pressure-Related Symptoms
The following symptoms are all related to high vacuum pressure. Each
symptom is discussed in more detail in the following pages.
The mass flow controller is calibrated for methane and the high vacuum gauge
controller is calibrated for nitrogen, so these measurements are not accurate
in absolute terms (Table 5). They are intended as a guide to typical observed
readings. They were taken with the following set of conditions:
Source temperature 300 °C
Quad temperature 150 °C
Interface temperature 280 °C to 320 °C
Helium carrier gas flow 1 mL/min
Table 5 Typical analyzer vacuum with reagent gas flow
Collision cell gas flow on Collision cell gas flow off
MFC
(%)
Rough vac High vac Rough vac High vac
10 1.77
× 10
–1
7.15 × 10
–5
1.33 × 10
–1
2.56 × 10
–6
15 1.86 × 10
–1
7.19 × 10
–5
1.43 × 10
–1
3.00 × 10
–6
20 1.94 × 10
–1
7.23 × 10
–5
1.53 × 10
–1
3.45 × 10
–6
25 2.02 × 10
–1
7.27 × 10
–5
1.63 × 10
–1
3.86 × 10
–6
30 2.10 × 10
–1
7.31 × 10
–5
1.71 × 10
–1
4.30 × 10
–6
35 2.18 × 10
–1
7.39 × 10
–5
1.80 × 10
–1
4.76 × 10
–6
40 2.25 × 10
–1
7.43 × 10
–5
1.88 × 10
–1
5.18 × 10
–6

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals