EasyManuals Logo

Agilent Technologies 7000 Triple Quadrupole GC/MS User Manual

Agilent Technologies 7000 Triple Quadrupole GC/MS
214 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
CI Troubleshooting 3
Troubleshooting and Maintenance Manual 59
Poor vacuum without reagent gas flow
Excess water in the background
Scan from 10 to 40 m/z. A large peak at m/z 19 (>m/z 17) indicates water in the
background. If water is present, allow the instrument to bake out more and
flow reagent gas through the lines to purge any accumulated water.
Air leak
Run Methane Pretune. See the Operation Manual. A visible peak at m/z 32
indicates air in the system. Check for and correct any leaks. See “Air Leaks” on
page 54.
The foreline pump is not working properly
For the standard foreline pump, replace the pump oil. If that does not help, or
for the dry foreline pump, it may be necessary to replace the pump. Contact
your local Agilent Technologies Customer Engineer.
The turbo pump is not working properly
Check the pump speed. It should be at least 95%. Contact your local Agilent
Technologies service representative.
CAUTION
Use of ammonia as reagent gas can shorten the life of the foreline pump oil (with
standard pump) and possibly of the foreline pump itself. See “To Minimize Foreline
Pump Damage from Ammonia” on page 124.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals