General Troubleshooting 2
Troubleshooting and Maintenance Manual 39
Foreline pressure is too low
If the pressures you observe are below 20 mTorr, check for the following:
Column (carrier gas) flow is too low
Column plugged or crushed by an overtightened nut
Collision gas flows are too low
Empty or insufficient carrier gas supply*
Bent or pinched carrier gas tubing*
Foreline gauge is not working correctly
* This could create a fault condition in the GC that would prevent the GC
from operating.
Analyzer chamber pressure is too low
If the pressures you observe are below 1 × 10
-6
Torr, check for the following:
Column (carrier gas) flow is too low
Collision gas flows are too low
Column plugged or crushed by overtightened nut
Empty or insufficient carrier gas supply*
Bent or pinched carrier gas tubing*
* This could create a fault condition in the GC that would prevent the GC
from operating.