General Troubleshooting 2
Troubleshooting and Maintenance Manual 35
High background
Pressure in the analyzer chamber is too high
Air leak
Contamination
Mass assignments are incorrect (scan mode only)
Small shape changes at the top of the mass peaks can cause 0.1 m/z shifts in
mass assignments. Shifts greater than 0.2 m/z indicate a possible malfunction.
MS has not had enough time to reach thermal equilibrium
Large variations in the temperature of the laboratory
MS has not been tuned recently or at the temperature at which it is
operating
Incorrect tune file (inappropriate parameters)
No voltage to extractor lens (if using extractor EI source)
Peaks have precursors
The tune report lists the size of the precursors for the tune masses. Small
precursors are not unusual. If the precursors are unacceptably large for your
application, one of the following may be responsible:
Repeller voltage is too high
Peaks are too wide
Incorrect DC polarity on the quadrupole mass filter
Dirty quadrupole mass filter
Peak widths are very low
MS has not had enough time to reach thermal equilibrium
Large variations in the temperature of the laboratory
Incorrect tuning
Calibration vial(s) empty or almost empty
Calibration valve(s) not working correctly
Dirty ion source