General Troubleshooting 2
Troubleshooting and Maintenance Manual 37
High mass sensitivity is poor
This refers to a condition where the absolute abundance at the upper end of
the mass range is poor. Absolute abundance should not be confused with the
relative abundance (percentage) of m/z 502 to m/z 69. Sensitivity at high
masses can be excellent even if the relative abundance of m/z 502 is low.
Wrong tune program
Wrong tune file
Repeller voltage is too low
Not enough time for the MS to warm up and pump down
Analyzer chamber pressure is too high
Column (carrier gas) flow is too high
Poor filament operation
Dirty ion source
Air leak
Incorrect DC polarity on the quadrupole mass filter
Collision cell gas flow is too high
Collision cell voltages are too high
Prefilter or postfilter
No voltage to the extractor lens (if using extractor EI source)