General Troubleshooting 2
Troubleshooting and Maintenance Manual 43
Difficulty with the HED supply
The only time this error occurs is if the output of the supply cannot get to its
destination (the HED).
Large peak, such as the solvent peak, eluted while the analyzer was on
Pressure in the analyzer chamber is too high
Detector is not working correctly
MS electronics are not working correctly
Difficulty with the high vacuum pump
This indicates the pump failed to reach 50% of full speed within 10 minutes or
experienced a fault.
You must switch the MS off and back on to remove this error message. Be sure
the turbo pump has slowed down before switching off the MS. The message
will reappear if the underlying problem has not been corrected.
Large vacuum leak is preventing the turbo pump from reaching 50% of full
speed
Foreline pump is not working correctly
Turbo pump is not working correctly
Turbo pump controller is not working correctly
MS electronics are not working correctly
High foreline pressure
Excessive carrier gas flow (typically > 5 mL/min)
Excessive solvent volume injected
Large vacuum leak
Severely degraded foreline pump oil
Collapsed or kinked foreline hose
Foreline pump is not working correctly
Internal MS communication fault
MS electronics are not working correctly

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals