CI Troubleshooting 3
Troubleshooting and Maintenance Manual 63
If MS was recently switched from EI to CI
Interface tip seal not installed
No reagent gas flow
Analyzer not sealed (big air leak)
Wrong tune file loaded or tune file corrupted
Ion source not assembled or connected correctly
Wrong reagent gas selected for the tune file (looking for wrong ions)
No PFDTD peaks in PCI
Incorrect reagent gas. There are no PCI PFDTD peaks created with
isobutane or ammonia. Switch to methane.
Analyzer not sealed (big air leak)
No calibrant in vial
Defective calibration valve(s)
Air leak in carrier or reagent gas path
No reagent gas peaks in NCI
Reagent gases do not ionize in NCI; look for background ions instead
Verify tune parameters
If no background ions are visible, go back to methane PCI
No PFDTD calibrant peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
)
Verify tune parameters
Go back to methane PCI
No sample peaks in NCI
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
)
Go back to methane PCI
Poor quality reagent gas (purity less than 99.99%)
Large peak at m/z 238 in NCI OFN spectrum
Look for background ions: 17 (OH
), 35 (Cl
), and 235 (ReO3
)
Find and fix your small air leak

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals