CI Troubleshooting 3
Troubleshooting and Maintenance Manual 71
Reagent gas ion ratio is difficult to adjust or unstable
The interface tip seal is incorrectly placed, damaged, or missing.
Inspect the interface tip seal. If necessary, remove and reinstall it to ensure a
good seal with the CI ion source. Replace it if it is damaged. Install it if it is
missing.
Residual air and water in the MS or in the reagent gas supply lines
Run the methane pretune. Air will appear as a peak at m/z 32 and excessive
water as a peak at m/z 19 > m/z 17. If either of these conditions is present,
purge the reagent gas supply lines and bake out the MS. Continued presence of
a large peak at m/z 32 may indicate an air leak. After correcting the problems,
you may need to clean the ion source.
Air leak
Run Methane Pretune (in PCI). Large peak at m/z 32 indicates air in the
system. Check for and correct any leaks. See “Air Leaks” on page 54.
The reagent gas supply is at the wrong pressure.
Check the regulator on the reagent gas supply. It should be adjusted to 20 psi
(140 kPa).
A leak in the reagent gas delivery path. This is especially likely if you have set
the methane flow much higher than normal and the ratio is still too low.
Check the reagent gas path. Tighten fittings.
The CI ion source is dirty.
Clean the ion source. See the Operation Manual for more information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7000 Triple Quadrupole GC/MS and is the answer not in the manual?

Agilent Technologies 7000 Triple Quadrupole GC/MS Specifications

General IconGeneral
BrandAgilent Technologies
Model7000 Triple Quadrupole GC/MS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals