EasyManua.ls Logo

Agilent Technologies 7010 Series - To Set the Mode for the Optional Jetclean System

Agilent Technologies 7010 Series
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
5 Operating in EI Mode
To Set the Mode for the Optional JetClean System
7000/7010 Series TQ GC/MS Operating Manual 113
To Set the Mode for the Optional JetClean
System
1 Click the MS Parameters icon in the Instrument Control view.
2 Click Source Cleaning from the Triple Quadrupole MS Method Editor
navigation panel.
3 Select a mode from the Operation dropdown menu. (See Figure 35.)
For detailed information on the JetClean system see the JetClean Operating
manual.
Figure 35. TQ MS Method Editor Source Cleaning panel

Table of Contents

Related product manuals