EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>7010 Series

Agilent Technologies 7010 Series User Manual

Agilent Technologies 7010 Series
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #86 background imageLoading...
Page #86 background image
4 Installing 7890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using a Standard Column Nut
86 7000/7010 Series TQ GC/MS Operating Manual
To Install a Capillary Column in the GC/MS
Interface Using a Standard Column Nut
This procedure is for the installation of a capillary column directly into the
analyzer. There are two types of column nuts that can be used in the GC/MS
interface: The standard column nut explained here, and the self-tightening
column nut explained in the previous section.
Materials needed
Interface tip seal (G3870-20542) (See Figure 20 and Figure 21 on page 83.)
Tip seal spring (G7005-20024)
Knurled tip seal retainer (G3870-20547)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Interface column nut (05988-20066)
Ferrules
0.3 mm id, for 0.10 mm id columns (5062-3507)
0.4 mm id, for 0.20 and 0.25 mm id columns (5062-3508)
0.5 mm id, for 0.32 mm id columns (5062-3506)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses
Wrench, open-end, 1/4-inch and 5/16-inch (8710-0510)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7010 Series and is the answer not in the manual?

Agilent Technologies 7010 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7010 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals