8CI Maintenance
To Clean the CI Source
232 7000/7010 Series TQ GC/MS Operating Manual
To Clean the CI Source
Materials needed
Abrasive paper (5061-5896)
Alumina abrasive powder, 100 g (393706201)
Aluminum foil, clean
Cloths, clean (05980-60051)
Cotton swabs (5080-5400)
Glass beakers, 500 mL
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Solvents
Acetone, reagent grade
Methanol, reagent grade
Methylene chloride, reagent grade
Ultrasonic bath
Preparation
1 Disassemble the CI source. (See “To Disassemble the CI Source” on
page 229.)
2 Collect the following parts to be cleaned for a CI source: (See Figure 78 on
page 233.)
Repeller
Source body
Drawout plate
Drawout cylinder
Ion focus lens
Entrance lens
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.