EasyManuals Logo

Agilent Technologies 7010 Series User Manual

Agilent Technologies 7010 Series
242 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
5 Operating in EI Mode
View the TQ parameters
7000/7010 Series TQ GC/MS Operating Manual 99
GC/MS Instrument Configuration program. See Configure the system in the
7890A+ or 7890B GC in the GC/MS Instrument Configuration program help.
9 In the GC/MS Instrument Configuration program, edit the TQ IP address
using these new parameters.
View the TQ parameters
This procedure allows you to monitor TQ parameters without running the
MassHunter GC/MS Acquisition software. It is useful when manually pumping
down the TQ and monitoring the instrument's status.
1 Press the MS On/Off switch to start the instrument.
2 Enter the instrument IP address into a PC web browser URL to display the
WUI Main Menu page. (See Figure 28.)
Figure 28. WUI main menu

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7010 Series and is the answer not in the manual?

Agilent Technologies 7010 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7010 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals