5 Operating in EI Mode
View the TQ parameters
7000/7010 Series TQ GC/MS Operating Manual 99
GC/MS Instrument Configuration program. See Configure the system in the
7890A+ or 7890B GC in the GC/MS Instrument Configuration program help.
9 In the GC/MS Instrument Configuration program, edit the TQ IP address
using these new parameters.
View the TQ parameters
This procedure allows you to monitor TQ parameters without running the
MassHunter GC/MS Acquisition software. It is useful when manually pumping
down the TQ and monitoring the instrument's status.
1 Press the MS On/Off switch to start the instrument.
2 Enter the instrument IP address into a PC web browser URL to display the
WUI Main Menu page. (See Figure 28.)
Figure 28. WUI main menu