4 Installing 7890 GC Columns
To Install a Capillary Column in the GC/MS Interface Using the Self-Tightening Column Nut
7000/7010 Series TQ GC/MS Operating Manual 81
To Install a Capillary Column in the GC/MS
Interface Using the Self-Tightening Column
Nut
This procedure is for the installation of a capillary column directly into the
analyzer using the Agilent recommended self-tightening column nut.
Materials needed
Interface tip seal (G3870-20542) (See Figure 20 and Figure 21 on page 83.)
Knurled tip seal retainer (G3870-20547)
Tip seal spring (G7005-20024)
Column cutter, ceramic (5181-8836) or diamond (5183-4620)
Flashlight
Magnifying loupe
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Self-tightening column nut for GC\MS interface (5190-5233)
Ferrules, Vespel
0.27 mm id, for 0.10 mm id columns (5062-3518)
0.37 mm id, for 0.20 mm id columns (5062-3516)
0.40 mm id, for 0.25 mm id columns (5181-3323)
0.5 mm id, for 0.32 mm id columns (5062-3514)
0.8 mm id, for 0.53 mm id columns (5062-3512)
Septum (may be old, used inlet septum)
Safety glasses