8CI Maintenance
To Remove the 7010 Series CI Source
7000/7010 Series TQ GC/MS Operating Manual 221
To Remove the 7010 Series CI Source
Materials needed
Gloves, clean, lint-free (Large 8650-0030) (Small 8650-0029)
Tweezers (8710-2460)
Procedure
1 Vent the MS. (See “To Vent the MS” on page 124.)
2 Open the analyzer chamber. (See “To Open the Front Analyzer Chamber” on
page 156.)
3 Disconnect the wires from the ion source. (See“To Connect or Disconnect
7010 Series CI Source Wiring” on page 217.)
4 Remove the two large thumbscrews that hold the ion source in place.
5 Pull the ion source out of the source radiator, and place it in its storage
container.
WARNING
The analyzers, GC/MS interface, and other components in the analyzer
chamber operate at very high temperatures. Do not touch any part until you
are sure it is cool.
CAUTION
Always wear clean gloves to prevent contamination when working in the
analyzer chamber.
CAUTION
Use an antistatic wrist strap, and take other antistatic precautions before
touching analyzer components.
CAUTION
When disconnecting leads, pull on the connectors, not on the wires.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7010 Series and is the answer not in the manual?

Agilent Technologies 7010 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7010 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals