1Introduction
The 7000/7010 Series TQ GC/MS
16 7000/7010 Series TQ GC/MS Operating Manual
The 7000/7010 Series TQ GC/MS
The 7000/7010 Series TQ GC/MS is a standalone capillary GC detector for use
with the Agilent 8890, 9000, or 7890 GCs. (See Figure 1 on page 18.) The TQ
GC/MS features:
One split flow turbomolecular vacuum pump
Rotary vane or optional dry scroll foreline pump
A choice of two independently MS-heated high sensitivity EI sources
An optional JetClean system for cleaning the ion source in place under
vacuum
CI and EI modes available (PCI/NCI/EI)
Two independently MS-heated hyperbolic Quads
Single hexapole CC
HED EM
Independently GC-heated GC/MS interface with interface tip seal
Independently GC-controlled CC gas flows
Physical description
The 7000/7010 Series TQ GC/MS is a rectangular box, approximately 47 cm
high, 35 cm wide, and 86 cm deep. The weight is 59 kg for the turbo pump
mainframe and 64 kg for the mainframe with CI. The attached foreline (roughing)
pump weighs an additional 22.2 kg.
The basic components of the instrument are:
Frame/cover assemblies
Vacuum system
GC/MS interface
Ion source
Electronics
CC
Detector
Front and rear analyzers

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7010 Series and is the answer not in the manual?

Agilent Technologies 7010 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model7010 Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals