Bath Heater Error
The Bath Error - Heater Disabled screen shows all of the possible errors
associated with the bath heater and the bath probe. A red X displays
before the current error condition. If a bath heater error is encountered,
the current running method is not stopped and you are given the option
of correcting the error and continuing the test. Press Retry to retest for
the error. If the error has been corrected, the dialog clears and you are
returned to the Main screen. Press Ignore to ignore the error until the
unit is powered on / off again.
Bath Heater Error
Action
Heater Probe Open or No Comm.
Ensure the bath heater cable is connected.
Check cable.
Heater Air Pocket or Heater OFF
The heater / circulator temperature is more than 6 °C hotter
than the bath temperature probe indicating an air pocket or
stagnant hot water in the heater element. Follow the
priming procedure on the back of the Wide Input Range
Heater / Circulator.
Bath Temperature 2 ºC Over Setpoint
Check for a faulty probe.
Contact Agilent Customer Care (“Obtaining Warranty and
Other Services” on page 134).
Bath Overtemperature
Turn off the bath heater.
Contact Agilent Customer Care (“Obtaining Warranty and
Other Services” on page 134).
Bath Probe Open
Ensure the bath probe is plugged in.
Replace the bath probe.
Bath Probe Short
Replace the bath probe.
Contact Agilent Customer Care (“Obtaining Warranty and
Other Services” on page 134).
Heater Probe Short
Contact Agilent Customer Care (“Obtaining Warranty and
Other Services” on page 134).
Table 1 7 Error Screens (Red)

Table of Contents

Other manuals for Agilent Technologies 708-DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 708-DS and is the answer not in the manual?

Agilent Technologies 708-DS Specifications

General IconGeneral
BrandAgilent Technologies
Model708-DS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals