72 708-DS / 709-DS Operator’s Manual
4 Operating the 708-DS / 709-DS
Method Properties Screen 3
Table 4 Method Properties Screen 2
Option Function
Enable Manifold Enabling this option activates the optionally installed sampling
manifold. The manifold lowers the sampling probes into the solution,
allowing automated sample collection as well as the temperature probes
for automated temperature measurement.
Initial Temp Enables recording and printing of the initial test temperature.
Final Temp Enables recording and printing of the final test temperature.
Manifold Down
Lead Time
This is the amount of time that the manifold lowers prior to the sample
time point. This allows for equilibration of the temperature probes
before vessel temperature measurement and for priming the system
when an external pump is connected.
Manifold Down
Duration
This is the amount of time that the manifold remains at its sampling
position. The sum of the Lead Time and Duration must be greater than
or equal to 30 seconds for accurate temperature readings.

Table of Contents

Other manuals for Agilent Technologies 708-DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 708-DS and is the answer not in the manual?

Agilent Technologies 708-DS Specifications

General IconGeneral
BrandAgilent Technologies
Model708-DS
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals