Logs
Three logs are accessible from the keypad: the run log, the
maintenance log, and the system event log. To access the
logs, press [Logs] to toggle to the desired log. The display
will indicate the number of entries the log contains. Scroll
through the list.
Sequence log
The sequence log is cleared at the start of each new
sequence. During the sequence, any deviations from the
planned methods (including keypad intervention) are listed
in the sequence log table.
Maintenance log
The maintenance log contains entries made by the system
when any of the user- defined component counters reach a
monitored limit. The log entry contains a description of the
counter, its current value, the monitored limits, and which of
its limits has been reached. In addition, each user task
related to the counter is recorded in the log, including
resetting, enabling or disabling monitoring, and changing
limits or units (cycles or duration).
Event log
The event log records significant events during HS operation.
Some of the events also appear in the sequence log if they
occur during sequence execution.

Table of Contents

Other manuals for Agilent Technologies 7697A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7697A and is the answer not in the manual?

Agilent Technologies 7697A Specifications

General IconGeneral
BrandAgilent Technologies
Model7697A
CategoryMedical Equipment
LanguageEnglish

Related product manuals