Sequences 7
Operation 65
When more than one consecutive entry exists for a vial,
and the entries use the same method, the HS will
puncture the vial once, then make the extractions and
injections. The HS will not perform multiple vial
punctures.
Extraction mode is Concentrated.
If a vial appears more than once in a sequence, or if the
number of injections is greater than one, the vial is
completely reprocessed for each entry or injection.
See also “To Create a Sequence” and “Sequences and
Throughput (111 vial model)”.
How many sequences can the HS store?
The HS can store up to 9 sequences internally. In addition,
the HS can temporarily store sequences that are downloaded
from the data system.