76 Operation
8 Running Samples
To Run a Series (Sequence) of Samples
1 Create or load the first method. (This prepares the HS
temperatures and flows for the first sample.)
If needed, create the method. See “To Create a
Method”.
If already created, load the existing method. See “To
Load a Method”.
2 Prepare the sample vials.
3 Load the sample vials into the tray. See “To Load a
Sample into the Tray (111 Vial Model)” or “To Load a
Sample into the Tray (12 Vial Model)”.
4 Create or load the sequence to run.
If needed, create the sequence. See “To Create a
Sequence”.
If already created, load the existing sequence. See “To
Load a Sequence”.
5 Press [Start].
When the HS becomes ready (all temperatures and flows
are at method setpoint), the HS will begin processing the
samples.
To view the progress of the vials, press [Status]. See also
“To View Vial and Sequence Status”.