68 Operation
7 Sequences
Sequences and Throughput (111 vial model)
The HS optimizes throughput by checking the temperature
and shaking parameters for the vials specified in the current
sequence. When consecutive vials share the same
temperature and shaking settings, the HS will examine the
timing parameters for the samples, then calculate the best
times to place each vial into the oven. This approach
maximizes the number of vials equilibrating at a time.
Vials without the same temperature and shaking parameters
will not be handled until the preceding samples leave the
oven.
For more information, see the Advanced Operation Guide.