EasyManuals Logo

Agilent Technologies 7820A Advanced User's Guide

Agilent Technologies 7820A
130 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #77 background imageLoading...
Page #77 background image
Detectors 5
Advanced User Guide 77
3 Verify that makeup gas type is the same as that plumbed to
your instrument (next to Makeup line in the parameter list).
Change the gas type, if necessary.
4 Set the reference gas flow rate.
5 If you are using packed columns, turn off the makeup gas (or
proceed to step 6 and enter 2 to 3 mL/min, see “TCD carrier,
reference, and makeup gas” on page 73) and proceed to step
7
6 If you are using capillary columns:, choose a flow mode and set
the makeup gas flow or combined flow.
7 Turn on the filament. Allow about 30 minutes for thermal
stabilization. A longer period may be needed for the highest
sensitivity.
8 If necessary, turn Negative polarity [On/Yes] to invert
negative-going peaks. When a sample contains components
giving both positive- and negative-going peaks, Negative
polarity can be switched on and off during a run as a
timetable event.
Example: Packed mode (packed and large capillary columns)
Column flow is 15 to 60 mL/min. Set the reference flow to 1.5
times the sum of column flow + makeup flow.
Makeup gas is recommended with all capillary columns. It
allows the column to be inserted all the way into the detector
and withdrawn 1 mm. If makeup is not used, the column must
be no more than 3 mm above the ferrule. Minimum makeup flow
is 1 mL/min.
1/8-inch stainless steel column If column flow is 30 mL/min,
set the reference flow to 30 × 1.5 = 45 mL/min. Total detector
flow is 30 + 45 = 75 mL/min.
10 m × 0.53 mm column If column flow is 15 mL/min and
makeup flow = 2 mL/min, set the reference flow to 1.5 × (15 + 2)
= 25.5 mL/min. Total detector flow is 17 + 25.5 = 42.5 mL/min.
Example: Capillary mode (small capillary columns)
If combined column plus makeup flow is between 5 and
10 mL/min, set the reference flow at 3× the combined flow. For
a combined flow between 10 and 15 mL/min, use a multiplier of
2. This will bring the TCD within 25% of the maximum response.
For further optimization, adjust the reference flow.

Table of Contents

Other manuals for Agilent Technologies 7820A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 7820A and is the answer not in the manual?

Agilent Technologies 7820A Specifications

General IconGeneral
BrandAgilent Technologies
Model7820A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals