EasyManuals Logo

Agilent Technologies 81101A User Manual

Agilent Technologies 81101A
94 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #42 background imageLoading...
Page #42 background image
42
Getting Started
Setting Up a Burst Signal
Programming
example
If you want to include this burst signal in your GP-IB program use the
following command lines. The comment lines starting with a # are not
required.
# Reset the instrument to start from a defined, default status.
*RST
# Switch off the automatic display update to increase programming
# speed.
:DISPlay OFF
# Set the instrument to burst mode by selecting a burst count
# of 2. Choose double pulses. Select triggered mode by selecting
# PLL as the trigger source and set the burst repetition to 2 µs.
:TRIGger:COUNt 2 # Set Burst mode with burst count of 2
:ARM:SOURce INT2 # Set Triggered mode with PLL as
# trigger source
:ARM:PERiod 2US # Set burst repetition of 2 µs
:PULSe:DOUBle ON # second channel generates double pulses
# per period
# Set the double pulse delay to 200 ns and the width to 100 ns.
:PULSe:PERiod 500NS
:PULSe:WIDTh 100NS
:PULSe:DOUBle:DELay 200NS
# Set the output amplitude to 2 Volts and the offset to
# 0.0 Volts.
:VOLTage 2V
# Enable the output.
:OUTPut ON

Other manuals for Agilent Technologies 81101A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 81101A and is the answer not in the manual?

Agilent Technologies 81101A Specifications

General IconGeneral
BrandAgilent Technologies
Model81101A
CategoryPulse Generator
LanguageEnglish

Related product manuals