EasyManua.ls Logo

Agilent Technologies 8163A - Error

Agilent Technologies 8163A
466 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error Messages Errors appearing on pop-up menus
390 Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition
Error
One or more of the following reasons may cause this error message:
firmware revision all
modus stepped sweep
parameter(s) lambda logging on
action switch lambda logging off
firmware revision <= 2.5x
modus continuous sweep
parameter(s) stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 10kHz)
lambda logging:
- coherence control active
- cycles 1
- trigger step finished
action check settings and correct where appropriate
firmware revision > 2.6
modus continuous sweep
parameter(s) stop wavelength <= start wavelength
sweep time too small (< 0.001s)
sweep time too large (t > 400s)
trigger frequency too high (f > 10kHz)
lambda logging:
- coherence control active
- trigger step finished
action check settings and correct where appropriate
All manuals and user guides at all-guides.com

Table of Contents

Other manuals for Agilent Technologies 8163A

Related product manuals