How to clean connectors Cleaning Instructions
Agilent 8163A/B, 8164A/B & 8166A/B Mainframes, Sixth Edition 435
An Alternative Procedure
A better, more gentle, but more expensive cleaning procedure is to use an
ultrasonic bath with isopropyl alcohol.
1 Hold the tip of the connector in the bath for at least three minutes.
2 Take a new, dry soft-tissue and remove the alcohol, dissolved sediment and
dust, by rubbing gently over the surface using a small circular movement.
3 Blow away any remaining lint with compressed air.