3-8 Chapter 3
Adjustments and Correction Constants
Source Pretune Default Correction Constants (Test 45)
Source Pretune Default Correction Constants (Test 45)
Analyzer warm-up time: 30 minutes.
This adjustment writes default correction constants for rudimentary phase lock pretuning
accuracy.
1. Press
.
2. Observe the analyzer for the results of this adjustment routine:
If the analyzer displays Pretune Def DONE, you have completed this procedure.
If the analyzer displays FAIL, refer to Chapter 7 , “Source Troubleshooting.”
Preset System
SERVICE MENU
TESTS
45 x1
EXECUTE TEST
YES