10-42 Chapter 10
Service Key Menus and Error Messages
GPIB Service Mnemonic Definitions
GPIB Service Mnemonic Definitions
All service routine keystrokes can be made through GPIB in one of the following
approaches:
sending equivalent remote GPIB commands. (Mnemonics have been documented
previously with the corresponding keystroke.)
invoking the System Menu (MENUSYST) and using the analyzer mnemonic (SOFTn),
where “n” represents the softkey number. (Softkeys are numbered 1 to 8 from top to
bottom.)
GPIB programming information is provided in the analyzer’s programmer’s guide.
Invoking Tests Remotely
Many tests require a response to the displayed prompts. Since bit 1 of the Event Status
Register B is set (bit 1 = service routine waiting) any time a service routine prompts the
user for an expected response, you can send an appropriate response using one of the
following techniques:
Read event status register B to reset the bit.
Enable bit 1 to interrupt (ESNB[D]). See “Status Reporting” in the analyzer’s
programmer’s guide.
Respond to the prompt with a TESRn command (see “Tests Menu” on page 10-4).
Symbol Conventions
[ ] An optional operand
D A numerical operand
< > A necessary appendage
| An either/or choice in appendages

Table of Contents

Other manuals for Agilent Technologies 8720ES

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8720ES and is the answer not in the manual?

Agilent Technologies 8720ES Specifications

General IconGeneral
BrandAgilent Technologies
Model8720ES
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals