9-2 Chapter 9
Accessories Troubleshooting
Accessories Troubleshooting
Accessories Troubleshooting
Use this procedure only if you have read Chapter 4 , “Start Troubleshooting Here.” Follow
the procedures in the order given, unless instructed otherwise.
Measurement failures can be divided into two categories:
• Failures which don't affect the normal functioning of the analyzer but render incorrect
measurement data.
• Failures which impede the normal functioning of the analyzer or prohibit the use of a
feature.
This chapter addresses the first category of failures which are usually caused by the
following:
• operator errors
• faulty calibration devices or connectors
• bad cables or adapters
• improper calibration techniques
These failures are checked using the following procedures:
• “Inspect the Accessories” on page 9-4
• “Inspect the Error Terms” on page 9-5