1-16
Troubleshooting the Analyzer
To troubleshoot the power supply
2 Check the A96 Primary Power Supply assembly’s voltages.
a Set the power switch to off ( O ) and disconnect the power cord
from the rear panel.
b Remove the top cover.
c Wait five minutes to allow time for the power supply capacitors to
discharge.
d Remove the power supply shield located on top of the analyzer
behind the display.
e Remove the primary power cable connected to A95 P350.