Performance Test
Using Recommended Test Equipment Using Other Test Equipment
Measurement
Uncertainty
Ratio
Measurement
Uncertainty
Ratio
Harmonic distortion
±0.9 dB
>10:1
Input capacitance
±2.5 pF
>10:1
Input resistance
510
>10:1
DC offset NA NA
Spurious signals NA NA
Noise NA NA
Cross talk
Source-to-input NA NA
Channel-to-channel
±0.6dB
>10:1
Anti-alias filter
<25 MHz
±0.145 dB
>10:1
>25 MHz
±0.225 dB
>10:1
Source amplitude accuracy
>26.98 dBm
±0.145 dB
>10:1
<26.98 dBm and >46.98 dBm
±0.150 dB
>10:1
<46.98 dBm
±0.210 dB
>10:1
Input rtn loss
±1.5 dB
6.3:1
Source rtn loss
±1.5 dB
6.3:1
Source distortion
10 kHz ±1.1 dB
8.4:1
>10 kHz
±1.5 dB
6.3:1
NA (not applicable) internal test
Verifying Specifications Agilent 89410A
Measurement Uncertainty
2-52

Table of Contents

Other manuals for Agilent Technologies 89410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 89410A and is the answer not in the manual?

Agilent Technologies 89410A Specifications

General IconGeneral
BrandAgilent Technologies
Model89410A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals