171
Chapter 6 Error Messages
Instrument Errors
4
6
112 Not able to perform requested operation
The requested operation is not valid for the instrument.
Example: FUNC 3,BIWIRE2 (the module in Slot 3 is not an N2260A).
113 Block name not exist
In the 3499A/B/C a maximum two blocks can be defined. The two defined
blocks can be read and written, etc. If you read or write a block that has
not been previously defined, this error occurs.
114 Block name already exist
The instrument has detected a defined block name, while you are to define
it once more.
115 Two Blocks already exist
The instrument has detected two defined block names while you are to
define them once more.
116 Channel number out of range
The specified channel number is invalid for the module in the selected
slot. The channel number has the form (@snn), where s is the slot number
and nn is the channel number.
Example:
ROUT:CLOSE (@156)
201 Scan list is empty
202 Scan initiated
203 Scan init ignored
204 Trig ignored
205 Hardware trigger too fast
206 Too many channels
207 Card in use
208 N2282A execution error
300 Unable to execute this command in local mode
501 RS232 data receiving error
502 Internal command error
503 RS232 only - unable to execute on GPIB
There are three commands which are allowd only with the RS-232
interface:
SYSTem:LOCal, SYSTem:REMote, SYSTem:RWLock.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 3499A and is the answer not in the manual?

Agilent Technologies Agilent 3499A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 3499A
CategoryControl Systems
LanguageEnglish