EasyManua.ls Logo

Agilent Technologies Agilent 3499A

Agilent Technologies Agilent 3499A
357 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
305
Chapter 9 Specifications
N2261A 40-Channel GP Relay Module
4
9
N2261A 40-Channel GP Relay Module
INPUT CHARACTERISTICS
Total Channels: 40
Maximum Voltage:
Terminal-Terminal or
Terminal-Chassis:
200 V, dc or ac rms
Maximum Current:
Per Channel:
Per Module:
1 A, dc or ac rms
20 A, dc or ac rms
Maximum Power:
Per Channel:
Per Module:
60 W dc; 62.5 VA ac
1200 W dc; 1250 VA ac
Thermal Offset:
Per Channel: < 3 µV
Initial Closed
Channel Resistance:
< 0.5
Relay Life:
Mechanical:
Electrical:
10
8
(at 36000 operations/hour)
5 x 10
5
(1A load)
Maximum Scan Rate:
80 Chans/sec
DC ISOLATION (with terminal block)
Open Channel,
Channel-Channel:
(with 1 channel closed)
< (40
o
C, 50% RH):
< (40
o
C, 80% RH):
> 10
10
> 10
9
Channel-Chassis:
(with 1 channel closed)
< (40
o
C, 50% RH):
< (40
o
C, 80% RH):
> 10
10
> 10
9

Table of Contents