Chapter 2 Front-Panel Operation
To Configure for External Trigger
2
To Configure for External Trigger
You can use one of two modules for an external trigger. The built-in rear
panel TRIG IN and TRIG OUT pair (on the control module) or the EI
(External Increment) and the CC (Channel Closed) pair on a 44474A
module. Only one pair can be used at a time. You can specify which pair
to use and whether a trigger out pulse is sent out when a relay is closed
during scanning operations. Refer to “Scanning” on page 85 for more
details.
1. Press the Menu key. The CONFIG annunciator lights up and the
display shows the first level menu.
2. Turn the knob to show the second menu item, “CONF EXT TRIG”.
3. Press Enter. The display shows the next level menu.
CARD PAIR
CONF EXT TRIG
TRIG SLOT 0