Chapter 9 Specifications
N2260A 40-Channel MUX Module
9
AC ISOLATION / PERFORMANCE
a
(without terminal block)
Capacitance:
(with 1 channel closed)
Open Channel, Channel-
Channel:
HI-LO:
Channel-Chassis:
< 7 pF (2-wire);
< 75 pF (2-wire);
< 150 pF (2-wire);
< 7 pF (dual 2-wire)
< 45 pF (dual 2-wire)
< 90 pF (dual 2-wire)
Insertion Loss:
(with 50 termination)
100 kHz:
1 MHz:
10 MHz:
< 0.10 dB
< 0.20 dB
< 1.50 dB
Crosstalk:
(with 50 termination)
100 kHz:
1 MHz:
10 MHz:
< -70 dB (2-wire);
< -50 dB (2-wire);
< -30 dB (2-wire);
< -40 dB (1-wire)
< -25 dB (1-wire)
NA (1-wire)
a. With chassis of all instruments connected, and with the Lo Terminal of the input connected to the
Lo Terminal of the output (either directly or via the 3499A/B/C switching channels).