Chapter 9 Specifications
N2266A 40-Channel MUX Module
9
AC CHARACTERISTICS
a
Bandwidth (-3dB):
40 MHz
Capacitance:
(with 1 channel closed)
(2-wire)
Open Channel, Channel-Channel:
Channel-Chassis:
HI-LO:
< 7 pF
< 140 pF
< 50 pF
Insertion Loss:
(with 50 termination)
100 kHz:
1 MHz:
10 MHz:
40 MHz:
0.2 dB
0.3 dB
2.0 dB
3.0 dB
Crosstalk:
(with 50 termination)
100 kHz:
1 MHz:
10 MHz:
-75 dB
-55 dB
-33 dB
a. With chassis of all instruments connected, and with the Lo Terminal of the input connected to the Lo Terminal of the
output (either directly or via the 3499A/B/C switching channels).