Chapter 9 Specifications
N2267A 8-Channel High Current GP Module
9
AC CHARACTERISTICS
a
Capacitance:
(with 1 channel closed)
Open Channel, Channel-Channel:
Channel-Chassis:
< 10 pF
< 10 pF
Insertion Loss:
(with 50 termination)
100 kHz:
1 MHz:
0.10 dB
0.20 dB
Crosstalk:
(with 50 termination)
100 kHz:
1 MHz:
-75 dB
-55 dB
Environmental Conditions
Operating Pollution
Degree:
2
Operating Altitude:
3000 meters (10,000 ft)
Measurement
Category:
I, 1500 Vpk transient, 500 V over voltage transient
Operating
Temperature:
0 - 55
o
C
Operating Humidity:
<80% RH (0
o
C to 40
o
C), non-condensing
a. With chassis of all instruments connected, and with Lo Terminal of the input connected to Lo Terminal of the output
(either directly or via the 3499A/B/C switching channels)