EasyManuals Logo

Agilent Technologies Agilent 3499A User Manual

Agilent Technologies Agilent 3499A
357 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #322 background imageLoading...
Page #322 background image
322
Chapter 9 Specifications
N2270A 10-Channel High Voltage MUX Module
9
AC CHARACTERISTICS
a
Capacitance:
(with 1 channel closed)
Open Channel, Channel-Channel,
Hi-Lo:
Channel-Chassis:
< 7 pF
< 50 pF
Insertion Loss:
(with 50
termination)
100 kHz:
1 MHz:
0.1 dB
0.2 dB
Crosstalk :
(with 50
termination)
100 kHz:
1 MHz:
-70 dB
-50 dB
Environmental Conditions
Operating Pollution Degree:
2
Operating Altitude:
3000 meters (10,000 ft)
Measurement Category:
I, 1500 Vpk transient, 500 V over voltage transient
Operating Temperature:
0-55
o
C
Operating Humidity:
<80% RH (0
o
C-40
o
C), non-condensing
a. With chassis of all instruments connected, and with Lo Terminal of the input connected to Lo Terminal of the output
(either directly or via the 3499A/B/C switching channels).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Agilent 3499A and is the answer not in the manual?

Agilent Technologies Agilent 3499A Specifications

General IconGeneral
BrandAgilent Technologies
ModelAgilent 3499A
CategoryControl Systems
LanguageEnglish