91
Chapter 4 Features and Functions
Scanning
4
4
•
Front-Panel Operation: Use the S.List menu to set the scan
configuration as shown below.
If you select TIMER for either arm source or trigger source, an
additional menu is shown that enables you to set the time in seconds.
•
Remote Interface Operation: The following code sets the scan sweep to
immediately start, the channel scan is advanced by the timer, and 10
sweeps of the scan list will occur. A 2 second delay is also set for
channel 101.
ARM:SOURce IMMediate
! Set arm source to IMM.
ARM:COUNt 10
! Set arm count to 10 times.
TRIGger:SOURce TIMer
! Set trigger source to TIMer.
TRIGger:TIMer 5 ! Set the trigger timer to 5
seconds.
ROUTe:CHANnel:DELAY 2,(@101)
! Set a 2 second time delay for
channel 101.
CONFIG SCAN
ARM: IMM
TRIG: IMM
00000.0000 SECS
Press S.List. Turn knob. Press Ente
Turn the knob to select the arm sour
Press Enter.
Use arrow keys and knob to
select the delay time. Press Enter.
Turn the knob to select the trigger
source. Press Enter.
DELAY TIME
Turn the knob to select DELAY TIM
Press Enter.
COUNT: 00001
Use arrow keys and knob to
select the arm count. Press Enter.
SET ALL
Turn the knob to select ALL channel
or individual channels. Press Enter
3499
Press S.List to exit the menu.