2 Power Meter Operation
88 E4418B Power Meter User’s Guide
Saving and Recalling Power Meter Configurations
To reduce repeated setup sequences, up to ten power meter configurations
can be stored in the power meter’s nonvolatile memory. The GPIB address
and command set, the data stored in the sensor calibration tables and the
zeroing and calibration data are not stored by a preset. The calibration
table selected is not affected.
Power meter configurations are saved and recalled with the key.
To save the current measurement setup:
1 Press .
2 Use the and keys to scroll through the displayed files. When
the required file is highlighted, press .
3 Press .
If you require to modify the name of a file:
1 Press .
2 Use the and keys to scroll through the displayed files. When
the required file is highlighted, press . The power meter
displays the file name in a pop- up window. Modify this name (see
below) until the desired name is displayed.
Use or to modify the character on which the cursor is
currently positioned.
Use or to move to other characters.
Use and as required.
3 To confirm your choice, press .
To recall a measurement setup:
1 Press .
2 Use the and keys to scroll through the displayed files. When
the required file is highlighted, press .
3 Press .
Save
Confirm
Edit Name
Insert Char
Delete Char
Enter
Recall
Confirm