Power Meter Operation 2
E4418B Power Meter User’s Guide 47
Figure 2-5 “Edit Cal Screen”
The power meter is shipped with a set of predefined sensor calibration tables.
The data in these tables is based on statistical averages for a range of Agilent
Technologies power sensors.
These power sensors are:
DEFAULT
1
HP 8481A
HP 8482A
2
HP 8483A
HP 8481D
HP 8485A
HP R8486A
HP Q8486A
HP R8486D
HP 8487A
There are also ten sensor calibration tables named CUSTOM_0 through
CUSTOM_9 which do not contain any data when the power meter is shipped
from the factory.
1 DEFAULT is a sensor calibration table in which the reference calibration factor and calibration factors are 100%.
This sensor calibration table can be used during the performance testing of the power meter.
2 The Agilent 8482B and Agilent 8482H power sensors use the same data as the Agilent 8482A.