Power Meter Operation 2
E4418B Power Meter User’s Guide 67
Step Detection
To reduce the filter settling time after a significant step in the measured
power, the filter can be set to re- initialise upon detection of a step
increase or decrease in the measured power. Step detection can be set in
both manual and automatic filter modes.
To set step detection:
1 Press , , .
2 Press the softkey to access the filter menu.
3 Use the softkey to enable or disable step detection.