Menu Reference 3
E4418B Power Meter User’s Guide 133
Press this softkey to enter a channel offset in the range of –100 dB to
+100
dB. The default value is 0 dB. Use the , , , and
keys to change the value. To confirm your choice, press . This offset
can be used to compensate for a loss or gain. The offset is applied to the
measured power before the result is displayed.
is
automatically set to “On” when a value is entered using . Refer
to “Setting Channel Offsets” on page 63 for further information.
(Agilent E- Series power sensors or N8480 Series power sensors
(excluding Option CFT) only)
Press this softkey to set the range in which you require the power meter
to make the power measurement or set the power meter to autoranging.
Choose either “UPPER”, “LOWER” or “AUTO” from the selection
available using the
and keys. The default is “AUTO”. Refer
to “Setting the Range” on page 79 for further information.
Press this softkey to access a menu which allows you to enter the upper
and lower measurement limits. Refer to
“Setting Measurement
Limits” on page 70 for further information.
Press this softkey to toggle between the test limits being on or off. The
default is “Off”.
Press this softkey to enter the upper measurement limit in the range
–150 dBm to 230 dBm. The default value is 90.00 dBm. Use the ,
, , and keys to change the value. To confirm your
choice, press the appropriate measurement units. must be
greater than .
Press this softkey to enter the lower measurement limit in the range
–150 dBm to 230 dBm. The default value is –90.00 dBm. Use the ,
, , and keys to change the value. To confirm your
choice, press the appropriate measurement units. must be less
than .
Press this softkey to toggle the duty cycle value on or off. The default is
“Off”. The duty cycle value is set using
.
Offset
dB
Offset Off On
Offset
Range
Limits
Limits Off On
Max
Max
Min
Min
Min
Max
Duty Cycle Off On
Duty Cycle