Error Messages 4
User’s Guide 153
–330 Self-test Failed;RAM Battery Fault
Refer to “RAM Battery” on page 98 if you require a description of the battery test.
–330 Self-test Failed;Calibrator Fault
Refer to “Calibrator” on page 99 if you require a description of the calibrator test.
–330 Self-test Failed;ROM Check Failed
Refer to “ROM Checksum” on page 98 if you require a description of the ROM Checksum test.
–330 Self-test Failed;RAM Check Failed
Refer to “RAM” on page 98 if you require a description of the RAM test.
–330 Self-test Failed;Display Assy. Fault
Refer to “Display” on page 99 if you require a description of the Display test.
–330 Self-test Failed;Confidence Check Fault
Refer to “Confidence Check” on page 96 if you require a description of this test.
–330 Self-test Failed;Serial Interface Fault
Refer to “Serial Interface” on page 99 if you require a description of this test.
–350 Queue overflow
The error queue is full and another error has occurred which could not be recorded.
–361 Parity error in program
The serial port receiver has detected a parity error and consequently, data integrity cannot be
guaranteed.
–362 Framing error in program
The serial port receiver has detected a framing error and consequently, data integrity cannot
be guaranteed.
–363 Input buffer overrun
The serial port receiver has been overrun and consequently, data has been lost.
–410 Query INTERRUPTED
A command was received which sends data to the output buffer, but the output buffer
contained data from a previous command (the previous data is not overwritten). The output
buffer is cleared when power has been off or after
*RST (reset) command has been
executed.