EasyManua.ls Logo

Agilent Technologies E4418B - Page 58

Agilent Technologies E4418B
199 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2 Power Meter Operation
34 E4418B Power Meter User’s Guide
Agilent 8485D Prior to the power meter being zeroed and calibrated, an Agilent 11708A
30 dB reference attenuator and an APC 3.5 (f) to 50 Ω (m) N-Type adapter
(08485-60005) should be connected between the power sensor and the
reference calibrator. Remove this attenuator from the power sensor input
before making measurements.
Agilent 8487A
Agilent N8487A
Agilent N8488A
This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) to connect to the power meter. Remove this attenuator
before making measurements.
Agilent 8487D Prior to the power meter being zeroed and calibrated, an Agilent 11708A
30
dB reference attenuator and an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) should be connected between the power sensor and the
reference calibrator. Remove this attenuator from the power sensor input
before making measurements.
Sensor Model Connection Requirements

Table of Contents

Other manuals for Agilent Technologies E4418B

Related product manuals