Power Meter Operation 2
E4418B Power Meter User’s Guide 49
5 To remove a sensor calibration table entry, use the , , , and
keys to select the entry, press . If you delete the frequency, the
equivalent calibration factor is also removed and vice versa.
6 When you have completed editing the sensor calibration table, press
to return to the “Sensor Tables” screen.
Delete
NOTE
You can only edit the table name, you cannot delete it.
Done
NOTE
Ensure that the frequency points you use cover the frequency range of the signals you
want to measure. If you measure a signal with a frequency outside the frequency range
defined in the sensor calibration table, the power meter uses the highest or lowest
frequency point in the sensor calibration table to calculate the calibration factor.