EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 27
Performance Tests
Instruction for Using the Performance Test Software
2. Performance Test
Instruction for Using the Performance Test Software
Setting system controller (USB/GPIB interface)
The test program needs a USB/GPIB interface to control the test equipment from the
E5061A/E5062A.
Follow these steps to set the USB/GPIB interface:
Step 1. Connect the USB port of the E5061A/E5062A to the USB/GPIB interface. The USB/GPIB
Interface Detected dialog box (Figure 2-1) appears.
NOTE Do not connect two or more USB/GPIB interfaces.
Figure 2-1 USB/GPIB Interface Detected dialog box
Step 2. Confirm that VISA Interface Name is set to GPIB0 (1 in Figure 2-1) and SICL Interface
Name is set to hpib7 (2 in Figure 2-1) and then click the
Accept button (3 in Figure 2-1). If
the setting is correct, the procedure is complete. If the setting is different, click the
Edit
button (4 in Figure 2-1).
Step 3. The USB to GPIB Configuration dialog box (Figure 2-2) appears. Make the setting
enclosed in the thick lines in Figure 2-2 (1 in Figure 2-2) according to the figure and then
click the
OK button (2 in Figure 2-2).

Table of Contents

Other manuals for Agilent Technologies ENA Series

Related product manuals