EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 63
Performance Tests
E5062A Performance Test Record (for Option 150/250)
2. Performance Test
Direction: S11 (Phase) (Option 150 only)
Crosstalk Test
Direction: S21
Direction: S12 (Options 250 only)
System Dynamic Range Test
For Serial Prefix : JP1KL or MY441
Direction: S21
Frequency range
[Hz]
IF bandwidth
[Hz]
Test limit [deg
rms]
Test result [deg
rms]
300 k - 1 M 3 k 0.05
1 M - 100 M 3 k 0.03
100 M - 1.5 G 3 k 0.03
1.5 G - 3 G 3 k 0.03
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -110
100 M - 1.5 G 10 -110
1.5 G - 3 G 10 -110
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -110
100 M - 1.5 G 10 -110
1.5 G - 3 G 10 -110
Frequency range [Hz] IF bandwidth [Hz] Test limit [dB] Test result [dB]
300 k - 100 M 10 -115
100 M - 1.5 G 10 -115
1.5 G - 3 G 10 -115
300 k - 100 M 3 k -90
100 M - 1.5 G 3 k -90
1.5 G - 3 G 3 k -90