EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
251 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 59
Performance Tests
E5062A Performance Test Record (for Option 150/250)
2. Performance Test
RF Output Level Linearity Test (@ Port1)
CW Frequency: 300 kHz (relative to 0 dBm reference)
CW Frequency: 50 MHz (relative to 0 dBm reference)
CW Frequency: 1.5 GHz (relative to 0 dBm reference)
3 G 1.0 0.16
Frequency [Hz] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 0.75 0.03
-2.5 0.75 0.03
2.5 0.75 0.03
5 0.75 0.03
7.5 0.75 0.03
10 0.75 0.03
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 0.75 0.03
-2.5 0.75 0.03
2.5 0.75 0.03
5 0.75 0.03
7.5 0.75 0.03
10 0.75 0.03
Power level [dBm] Test limit [dB] Test result [dB] Measurement
uncertainty [dB]
-5 0.75 0.03
-2.5 0.75 0.03
2.5 0.75 0.03
5 0.75 0.03
7.5 0.75 0.03

Table of Contents